Home → DC100/200 ILT Radiometer Retrofit → Preparation and Tools Needed → Machine Preparation
1.2. Machine Preparation
Before moving on, make sure everything here has been completed.
- Remove any solution from the machine
- Remove funnels or C-Tubes from machine (depending on configuration)
- Make sure you have plenty of space to work around the machine
- Manually move the radiometer arm to Position 4
- Log into the machine as an Engineer
- Go to the Radiometer screen
- Under 'Radiometer Controls', select 'Mode: Jog to Sensor'
- Press 'Jog Up' until you get to position 4
- Once the arm is in position 4, go to the Main screen
- Press the 'About' button
- Press the 'Quit' button to exit the software
- From the Start menu, shut down the PC
- Turn off power to the machine by turning the main disconnect handle off