Home → DC100/200 ILT Radiometer Retrofit → Preparation and Tools Needed → Machine Preparation

1.2. Machine Preparation

Before moving on, make sure everything here has been completed.

 

  1. Remove any solution from the machine
  2. Remove funnels or C-Tubes from machine (depending on configuration)
  3. Make sure you have plenty of space to work around the machine
  4. Manually move the radiometer arm to Position 4
    • Log into the machine as an Engineer
    • Go to the Radiometer screen
    • Under 'Radiometer Controls', select 'Mode: Jog to Sensor'
    • Press 'Jog Up' until you get to position 4
  5. Once the arm is in position 4, go to the Main screen
  6. Press the 'About' button
  7. Press the 'Quit' button to exit the software
  8. From the Start menu, shut down the PC
  9. Turn off power to the machine by turning the main disconnect handle off

 

 