Home → DC100/200 ILT Radiometer Retrofit → Remove Existing Radiometer and Cabling → Begin Existing Probe Cable Removal

3.1. Begin Existing Probe Cable Removal

  1. Disconnect probe cables from probes

  2. Cut all zip ties on radiometer carriage

  3. Unfasten both ends of Igus bracket using a 1/16" Allen key

  4. Remove the Igus chain cover by ‘unzipping’ the entire length of the chain. Set cover aside.

  5. Ends of Igus will need to be removed to get radiometer cables through (use a flathead screwdriver to pry the ends apart)

  6. Cut all cable ties from the end of the Igus chain up to the roof of the machine

  7. Undo any zip ties around cable bundles secured to frame (path shown below)

  8. Remove radiometer cables back to the top of the stack light

  9. Remove inside front door skin (with buttons)

  10. Remove bulkhead on cable opening

  11. Cut any zip ties between radiometer and bulkhead on front button panel (be sure to pull back sleeving over cable bundle and cut any zip ties found inside sleeving