Home → DC100/200 ILT Radiometer Retrofit → Software Configuration → Assigning COM Ports
6.1. Assigning COM Ports
-
On the PC, open Device Manager
-
Expand ‘Ports (COM & LPT), you should see several communication ports, as well as 2 USB serial ports.
-
Unplug CBL30013 from the PC, one of the serial ports should disappear, whichever one stays on the screen is CBL30014 – Radiometer A – record this somewhere
-
Plug CBL30013 back in and unplug CBL300014, whichever one stays on the screen is CBL30013 - Radiometer B - record this somewhere
-
Go to the C://DC/Source/Source/Data folder and open ‘Serial Radiometer.xml’
-
Change the COM ports for radiometer A and B to the COM ports just recorded
-
Save the SerialRadiometer.xml file
-
Open the DC200 software and log in as Engineer
-
Go to the Settings > Config page and find the ‘Radiometers Installed’ setting
-
Set this to ‘ILT x 1’ or ‘ILTx2’ based on the number of ILT1000s in use
-
Press Save
-
Exit the software
-
Restart the software
-
Confirm that there are numbers displaying in the ‘Live Readings’ window(s) on the Radiometer screen